Lowest Measured Temperature 75° F / 25° C
Highest Measured Temperature 6500° F / 3500° C
Operating Wavelengths

Nine different wave bands available
0.7 to 1.0

0.91 to .97
0.7 to 1.08 and 1.08
1.0 to 1.2
1.64
2.1 to 2.4
3.43
4.87 to 5.3
7.92

 

The Ircon MODLINE 3 Thermometer is a non-contact, Infrared Radiation Thermometer designed to meet the critical needs of today's industrial requirements. The MODLINE 3 Series features a broad line of infrared thermometers to match specific applications in many industries. The system offers you the choice of temperature sensors with sophisticated single lens reflex (SLR) focusing that can view spots as small as 0.012 inches (0.3mm), or infrared sensors using two color (ratio) technology to tolerate more than 95% reduction in radiant intensity, or fiber optic sensors for difficult to reach or obstructed targets.

The MODLINE 3 Indicator/Processor consists of a high performance sensor and microprocessor-based digital Indicator/Processor. Many standard features are included and only a two-point On/Off controller/alarm or a PID controller are optional. Additionally, Ircon offers standard accessories to meet the demanding environmental requirements you may have, such as air purge and water cooling, sight tubes and mounting assemblies.

MODLINE 3 utilizes advanced microprocessor technology to process highly accurate temperature signals received from the temperature sensor. This data is accurately displayed on a highly visible LED digital display in °F or °C. Programming features and control functions are viewed on a bright, highly readable Alpha Numeric, Vacuum Fluorescent display.
  • Broad Line Microprocessor Based Electronics
  • Choice of SLR or Fiber Optics Temperature Sensing Heads
  • Choice of 9 Different Wavebands
  • Low Temperature Ratio (Two Color)
  • High Temperature Ratio (Two Color)
  • Patented, Focusable Fiber Optics
  • Large LED Temperature Display
  • On/Off or PID Control Option
BASIC SYSTEM - A MODLINE 3 consists of the Temperature Sensor, Panel Mounting Indicator Unit, 12 foot or 4 meter Interconnecting Cable, and one detailed Operations Manual. Complete specifications and descriptions are shown in Catalog MO-300

MODLINE 3 Sensor Selection Chart

Series
Wavelength
Temp Span
Applications
Series: 200
Wavelength: 0.7 - 1.0µ
900-4800 °F
500-2600 °C
Forging, annealing, hardening, foundries, most iron and steel. Incadescent processes including semiconductor processes.
Series: 200 Fiber Optic
Wavelength: 0.7 - 1.0µ
1200-4800 °F
650-2600 °C
Series: 340
Wavelength: 3.43µ
75-1500 °F
25-800 °C
Thin film of polyethylene, polypropylene, vinyls and nylon.
Series: 600
Wavelength: 2.0 - 2.6µ
150 - 1400 °F
80-800 °C
Medium temperature applications for thick plastics, rubber, textiles and metals.
Series: 700
Wavelength: 4.8 - 5.3µ
100-2500 °F
50-2500 °C
Glass surface temperature in forming, bending, tempering, annealing and sealing.
Series: 800
Wavelength: 7.92µ
75-800 °F
25-400 °C
Thin films of polyester (PET) and fluorocarbon plastics, as well as ceramics and paint.
Series: 3G
Wavelength: 1.5 - 1.6µ
500-2500 °F
250-1400 °C
Medium temperature applications particularly ferrous and non-ferrous metals.
Series: 3G Fiber Optic
Wavelength: 1.5 - 1.6µ
650-2500 °F
350-1400 °C
Series: 3L
Wavelength: 1.00 - 1.20 or 1.5 - 1.6 and 1.08µ
500-1800 °F
250-1000 °C
Low temperature non-ferrous metals including aluminum using two-color ratio technique (Rolling Mills).
Series: 3R
Wavelength: 0.7 - 1.08 and 1.08µ
1300-6500 °F
700-3500 °C
For difficult high temperature processes such as molten metal, wire rod, vacuum furnaces and kilns, using two-color ratio technique.
Series: 3R Fiber Optic
Wavelength: 0.7 - 1.08 and 1.08µ
1300-6500 °F
700-3500 °C
Series: 3V
Wavelength: 0.91 - 0.97µ
400-1200 °C Specifically designed for silicon and gallium arsenide wafer temperature in MBE, MOCVD and MOMBE applications.
Series: 3V Fiber Optic
Wavelength: 0.91 - 0.97µ
500-1500 °C

 

MODLINE 3 Indicator/Processor
The MODLINE 3 Indicator/Processor is a compact, self-contained, microprocessor based signal processor and display designed for simple installation, operation and service.

The instrument is designed to operate with all MODLINE 3 temperature sensors and provides the input power to operate the sensor. Temperature signals from the sensor are displayed on a large visible, 0.8 inch (20.5 mm), LED display in °F or °C. MODLINE 3 offers you three standard selectable isolated analog output signals; 4 to 20mA, 0 to 20mA and 5µA per degree for your analog process control instruments. The analog outputs are adjustable to cover any portion of the instruments available temperature span to a minimum of 50° C or F. A standard RS-485 digital signal output is available for two-way digital communications for your digital process control instruments.

MODLINE 3 provides as standard features a complete range of signal conditioning functions, including Emissivity/E Slope, Track and Hold, Peak Picker and Response Time. Optional features offered are On/Off Controller or PID Controller.

The MODLINE3 Indicator/Processor includes a large LED display dedicated for temperature display and a smaller alpha numeric vacuum fluorescent digital display that is dedicated for displaying menu and control function inputs. Set-up and operation are easily configured using a four button keypad.

TEMPERATURE SENSORS

The MODLINE 3 system offers you the widest choice of non-contact infrared sensors to match any process. MODLINE 3 sensors offer sophisticated single lens reflex (SLR) focusing, that can view spots as small as 0.012 inches (0.3 mm); or for difficult, high temperature and harsh environments, some model series incorporate fiber optics for added flexibility. These rugged, high performance infrared sensors feature through-the-lens sighting of target, incorporating a circular reticle visible through the eyepiece that defines the exact spot to be measured.

The optical system, detector and electronics are housed in a rugged sealed casting rated NEMA 4 (IP66), to perform in the harsh industrial environments. MODLINE 3 sensors are designed to match existing mounting configurations and offer a choice of cooling and environmental protection accessories.

Fiber Optic Sensor Option
Our fiber optic temperature sensors are the ultimate in noncontact temperature measurement for difficult-to-read or obstructed targets. The following series are available as a fiber optic version:

SERIES & TEMPERATURE RANGE

200 Series 900 to 4800°F (650 to 2600°C)
3G Series 650 to 2500°F (350 to 1400°C)
3R Series 1300 to 6500°F (700 to 3500°C)
3V Series 500 to 1500°C
600 Series 150 to 1400°F (80 to 800°C)
340 Series 75 to 1500°F (25 to 800°C)
700 Series 100 to 2500°F (50 to 2500°C)
800 Series 75 to 800°F (25 to 400°C)

SPECIFICATIONS

CE Directives:
Meets EMC Directive 89/336/EEC for EMI/RFI and Safety Compliance
Tested to:
EN 50081-2: Generic Emissions Standard
EN 50082-2: Generic Immunity Standard
EN 61010-1: Safety Standard

Spectral Response
Each MODLINE 3 Series is designed for operation in a discrete wavelength band as shown in the Model Selection Chart

Calibration / Blackbody Accuracy @25°C:
For 200, 3G, 3R, and 3V Series within 0.6% of reading(+ 1 digit) or 5°F (3°C), whichever is greater.
For 340, 600, 700, 800 and 3L Series within 0.6% F.S. (+ 1 digit) or 5°F (3°C) whichever is greater.

Repeatability @25°C:
Within 0.1% of full-scale temperature (+1 digit)

Response Time At Analog Output
Adjustable from 0.01 to 60 seconds.

NEMA 4
The Temperature Sensor Housing and Indicator/Processor Enclosure are rated NEMA 4 (IP66).

OPERATING AMBIENT
without auxiliary cooling
Sensor* Series 200 and 3 G: 32 to 150°F (0 to 60°C)
Sensor* Series 3L and 3R: 32 to 130°F (0 to 55°C)
Sensor* Series 3V: 50 to 113°F (10 to 45°C)
Sensor* Series 340, 600, 700, and 800: 50 to 130°F (10 to 55°C)
Temperature Sensor Cable PVC: 220°F (100°C) Max. Temp.
Temperature Sensor Cable Silicone Rubber: 400°F (200°C) Max. Temp.
Indicator / Processor: 32 to 122°F (0 to 50°C)

Humidity
10 to 90% non-condensing

Enclosure
The Enclosure allows mounting two Indicator/Processors side-by-side in a 19 inch rack. Size: 7.56" x 5.67" (192 mm x 144 mm)

Selectable Analog Outputs

  • 0 to 20 mAdc with a 1000 Ohm max.
  • 4 to 20 mAdc with a 1000 Ohm max.
  • 5 uAdc / degree F or C ,(1mV/ ° [max 4000 mV]across 200 Ohm resistor, typical deviation 2° from display)
Digital Communication
RS-485 Digital Interface

Emissivity Range*
Emissivity is 0.100 to 1.000.

E Slope Range*
E Slope is 0.850 to 1.150.

Power Requirements
90 to 250 Vac, 50/60 Hz, 40VA max.

Weight
Temperature Sensor: 7 to 8 lb (3.2 to 3.6 kg)
Indicator/Processor: Approximately 9 lbs (4 kg)

UNIQUE TO MODLINE 3 SERIES 3L & 3R TWO COLOR (RATIO) THERMOMETER

3L Spectral Response
Adjacent bands: 1.50 to 1.60µ and 1.65 to1.71µ or 1.00 to 1.20µ and 1.65 to 1.71µ.

3R Spectral Response
Adjacent bands:0.7 to 1.08µ and a narrow band centered at 1.08µ.

Signal Reduction Range
Caused by low emissivity, non-resolved or an obliterated target above 575°F (300°C) for 3L series and above 1500°F (800°C) for 3R series. It will tolerate about 95% reduction in radiation intensity.

Invalid Alarm
Form C relay outputs rated at
24 V AC/DC;
@ 1 Amp. Resistive or Inductive.

PEAK PICKER

Rise time:
Same as selected Response Time from 0.01 to 60 seconds.

Decay Rate:
Adjustable 0.00 to 300°F per second or 0.00 to 300°C per second; depending on the instrument model

Peak Picker Reset:
Internal Reset has two Modes: Manual and Auto
External Reset via contact closure

ON/OFF CONTROLLER/ALARM WITH RELAY OUTPUTS

Response Time:
Same as selected Response Time from 0.01 to 60 seconds

Internal Set Point:
Adjustable over full temperature span of the instrument model

Remote Set Point:
Set Point 1 has an Input Signal of 4 to 20 mA, daisy chain compatible.

Relay Outputs:
Two Form C relay outputs rated at 24 V AC/DC;
@ 1 Amp. Resistive or Inductive

Relay Delay Time:
1 to 2 milliseconds

Hystersis:
2°F(2°C)

PID CONTROLLER WITH AUTO TUNE AND DEVIATION ALARMS

PID Output:
4 to 20 mAdc or 0 to 20 mAdc
1000 Ohm maximum

Remote Set Point:
Input Signal of 4 to 20 mA, daisy chain compatible

Internal Set Point:
Adjustable over full temperature span of the instrument model

Proportional Band (P):
0.1 to 200.0%

Reset Rate (I):
0.01 to 99.00 repeats per minute, with an Off position

Rate Time (D):
0.001 to 9.900 minutes, with an
Off position

Load Demand:
0 to 100 % of output

Manual Adjust:
0.0 to 100.0% of output

Deviation Alarm Set Points:
Adjustable for HI and LO set points over the full temperature span of the instrument model, each with an Off position

Deviation Alarm Output:
Relay Outputs for HI and LO Alarms use two Form C relay outputs rated at 24 V AC/DC;
@ 1 Amp. Resistive or Inductive.

Note: Continuous product improvements may change these specifications without notice.