The Mirage FO spot pyrometer consists
of one Fiber Optic infrared Sensor with Fiber Optic Cable,
Focusable Reimaging Lens assembly, one indicator unit, one 12
foot or 4 meter interconnecting cable and one detailed operations
manual. Complete description and specifications for Mirage
Fiber Optics in PB-FOMR and indication in Catalog MR-05.
The Fiber Optic Mirage Thermometer is available in three wavelengths
including a high temperature ratio model. These instruments
measure temperatures from 350 to 6500°F (180 to
3500°C).
Standard lens can measure objects at distances from
10" to infinity (250 mm to infinity), an optimal close-up
lens of 6" to 10" (150 to 250 mm) is also available.
The focusable lens and fiber optic cable is designed
to operate in ambient temperatures up to 400°F
(200°C) without auxiliary cooling, an optional;
extension tip can be used up to 575°F (300°C).
SPECIFICATIONS
FIBER OPTIC TEMPERATURE SENSOR AND INDICATOR/PROCESSOR
Spectral Response
Each MIRAGE Series Thermometer is designed for operation in a
discrete wavelength band. Refer to "Ordering Information
Selector Chart" for operating wavelengths for each
Series
Calibration Accuracy
Within 1% of full scale temperature or 10°F (6°C),
whichever is greater
Repeatability
Within 0.3% of full scale temperature ±1°
Response Time at Output:
Adjustable from 10 msec. to 10 sec. depending on Series
Ambient Temperature Range
Fiber Optic Sensor:
20 Series: 0° to 200°F (-17° to 90°C)
60 Series: 50° to 130°F (10° to 55°C)
OR Series: 30° to 130°F (0° to 55°C)
Fiber Optic Cable and Focusable Reimaging Lens:
0° to 400°F (-17° to 200°C); For 60
Series 300°F (150°C)
Extension Tip: 0° to 575°F (-17° to 300°C)
Indicator/Processor (All Series): 30° to 125°F
(0° to 50°C)
Relative Humidity:
10 to 90% non-condensing
Optical Characteristics
Reimaging Lens
Focal Range:10" to infinity (250 mm to infinity) standard
6" to 10" (150 to 250 mm) optional
Optical Resolution:D/30, D/60, D/30 x D150 per model selection
Extension Tip
Focal Range: Fixed focus
Optical Resolution:D/3
Analog Outputs
0 to 10 Vdc non-isolated (3 mA max.)
4 to 20 mA or 0 to 20 mA non-isolated (500 ohms load
max.)
(See Catalog MR-O5, for optional outputs)
Emissivity Control Range
Adjustable from 0.1 to 1.0
Power Requirements
100,115, 200 or 230 Vac ±10%, 50/60 Hz
Weight
Temperature Sensor: 7 to 8 lbs. (3.2 to 3.6 kg)
Indicator/Processor:Under 7 Ibs. (Under 3.2 kg)
Construction
Temperature Sensor: NEMA 4 (IP66)
Indicator/Processor:NEMA 12K (lP52)
UNIQUE TO "OR" SERIES TWO COLOR THERMOMETER
Spectral Response
Two adjacent bands: 0.7 to 1.08 microns narrow
band centered at 1.08 microns
E-Slope Adjustment
Adjustable E-slope from 0.85 to 1.15 with Greybody
switch position
Signal Reduction Range
(Caused by low emissivity, non-resolved or obliterated
target)
Above 1500°F (800°C) will tolerate approximately
95% reduction in radiation intensity
Invalid Alarm
Relay Contact ratings:
1A at 24 Vdc
0.5A at 120Vac, non-inductive load