IRCON's Fiber Optic InfraRail GT Infrared Thermometer offers
the powerful features and affordable price of the standard InfraRail
Infrared Thermometer, along with the flexibility and
ruggedness provided by an temperature sensor using fiber
optics. The InfraRail GT's unique focusable reimaging
lens sensor assembly and fiber optic cable offer the opportunity
to sight on targets where direct sighting is impossible
or, as in the case of glass furnaces, where extremely
high ambient temperatures are present.
The Fiber Optic InfraRail GT is used to measure temperatures
of the crown, side walls, breast wall, checkers and
molten glass temperature in the furnace, forehearth,
and runners.
- Designed Specifically For Temperature Measurement in Glass
Furnace Applications
- Miniature Infrared Sensing Head
- Quick Release Mount
- Patented, Focusable Fiber Optics Reimaging Lens
- 25 ft. (7.6m) Fiber Optic Cable
- Special Mounting Devices For Glass Furnace Applications
- 4-20 mA Two-Wire Signal Conditioner/Transmitter
- Utilizes Very Stable Silicon Detector
APPLICATIONS - Glass temperature measurement in the furnace, forehearth,
and runners.
Temperatures of the breast wall, checkers, bridge
wall, and crown.
INFRARAIL GT INFRARED THERMOMETER COMPONENTS :
The Fiber Optic InfraRail GT assembly consists of:
- InfraRail DIN-rail mounted Signal Processor with
response time adjustment, digital emissivity setting, peak picker,
and 4 to 20 mA linear output
- Fiber Optic Infrared Sensor
- 25ft (7.6m) of Fiber Optic Cable
- Focusable Reimaging Lens
- Model AAQ-1 Quick-Release Air Purge Assembly
- Model MB-2 Angle Mounting Bracket
- Model SI-12 Inconel Sight Tube (12in.[305mm])
Optional accessories may be purchased to replace those
supplied with the InfraRail GT Infrared Thermometer:
- Model SI-24 Inconel Sight Tube (12in.[610mm])
The InfraRail IR sensor operates at a 0.7 to 1.0 µm
wavelength. The sensor is a rugged, inherently stable
silicon photodiode with proven performance for over
40 years on glass furnace installations. The signal
processor can be mounted up to 1000 ft (300 m) from
the sensor and requires a simple two-wire shielded cable
for transmitting the sensor signal to the indicator.
The Reimaging Lens has a focal range of 10 in. to infinity
(254 mm to infinity). Most components are interchangeable
for easy routine maintenance. For more details, refer
to InfraRail Catalog DN-08
AMBIENT SPECIFICATIONS:
Temperature Limit
Inconel Sight Tube 2000°F (1093°C)
Focusable Reimaging Lens 400°F (200°C)
Fiber Optic Cable 400°F (200°C)
Fiber Optic Sensor 200°F (95°C)
Temperature Range and Resolution
Model GT-101 1500 to 2500°F D/60
Model GT-102 1800 to 3200°F D/60
Model GT-103 800 to 1300°C D/60
Model GT-104 900 to 1600°C D/60
Model GT-105 1000 to 1400°C D/60
Special temperature ranges are available on request.