Emissivity Values for Common Materials & Non-Metals - ircon.com
  Literature & Downloads   Ircon Advantage   IR Education   Contact Us   Change Language
 
Advanced Search  
  Home   Industry Applications   Products   Where To Buy
Home > IR Education > Emissivity Values - Common Materials

Emissivity Values for Common Materials & Non-Metals

 

Emissivity is the measure of an object's ability to emit infrared energy. Emitted energy indicates the temperature of the object. Emissivity can have a value from 0 (shiny mirror) to 1.0 (blackbody). Most organic, painted, or oxidized surfaces have emissivity values close to 0.95. The majority of Ircon sensors have an adjustable emissivity feature to ensure accuracy when measuring other materials, such as shiny metals.

Please note, these values are to be used as a guide only, as emissivity changes depending on the actual material surface and conditions.

Material Emissivity Values
1.0 µm 5.0 µm 7.9 µm 8-14 µm
Asbestos 0.9 0.9 0.95 0.95
Asphalt n.r. 0.9 0.95 0.95
Basalt n.r. 0.7 0.7 0.7
Carbon
  Unoxidized 0.8-0.95 0.8-0.9 0.8-0.9 0.8-0.9
Graphite 0.8-0.9 0.7-0.9 0.7-0.8 0.7-0.8
Carborundum n.r. 0.9 0.9 0.9
Ceramic 0.4 0.85-0.95 0.95 0.95
Clay n.r. 0.85-0.95 0.95 0.95
Concrete 0.65 0.9 0.95 0.95
Cloth n.r. 0.95 0.95 0.95
Glass
  Plate n.r. 0.98 0.85 0.85
  Gob n.r. 0.9 n.r. n.r.
Gravel n.r. 0.95 0.95 0.95
Gypsum n.r. 0.4-0.97 0.8-0.95 0.8-0.95
Ice n.r. 0.98 0.98
Limestone n.r. 0.4-0.98 0.98 0.98
Paint (non-Al.) 0.9-0.95 0.9-0.95
Paper (any color) n.r. 0.95 0.95 0.95
Plastic
  Opaque n.r. 0.95 0.95 0.95
  Over 20 mils n.r.
Rubber n.r. 0.9 0.95 0.95
Sand n.r. 0.9 0.9 0.9
Snow n.r. 0.9 0.9
Soil n.r. 0.9-0.98 0.9-0.98
Water n.r. 0.93 0.93
Wood (natural) n.r. 0.9-0.95 0.9-0.95 0.9-0.95

n.r. = not recommended

To optimize surface temperature measurement accuracy:
1. Determine the object emissivity for the spectral range of the instrument to be used for the measurement.
2. Avoid reflections by shielding object from surrounding high temperature sources.
3. For higher temperature objects, use shorter wavelength instruments, whenever possible.
4. For semi-transparent materials, such as plastic film and glass, assure that the background is uniform and lower in temperature than the object.

 
 
  Follow us on Twitter   Ircon YouTube Channel   Follow us on LinkedIn  
 
Fluke Process Instruments | About Ircon | Partner Login
Site Map | Terms of Use | Privacy Policy